Mail Support
labchemindia@yahoo.com
Call Support
03323633694 / 8910165145
Follow Us :

Product Details

Home / YAKOS 65 BIOSAFETY / SAFZONE A2 TYPE BIOSAFETY CABINET

SAFZONE A2 TYPE BIOSAFETY CABINET

HEPA FILTER: The clean H14 (EN 1822), 99.99% efficiency at 0.3 microns for supply and exhaust HEPA FILTER are used in safe zone biosafety cabinets to provide ISO CLASS 4 clean air to the work surface and remove 0.3-micron particles before entering the environment.

LIGHTING: An average light intensity of more than a thousand lux at the work surface.

Germicidal UV light: inbuilt UV light provides a minimum average intensity of 40 microwatts per centimetre square at the cabinet work surface

WORK AREA: SS304 single-piece stainless steel work tray

VIRO-CUTTM Anti-Microbial Stainless Steel is used in safe zone biosafety cabinets in the…. And water surface for the work area. This results in efficient contact cleaning for the exposure of bacteria in the stainless steel surface (option)

ECM BRUSHLESS DC MOTOR

Energy, efficient, less heat, no manned motors speed control, high torque and superior efficiency.

NIGHT IDLE MODE

To reduce Motor speed to 30%. This ensures the cabinet at a sterile working condition and saves significant electricity bills.

PATENTED HEPA FILTER

Replacement only one man and one minute to complete the filter changing without any tools. The class H14 per EW 1822 supply and exhaust paper filters are used in SafeZone. The cabinets allow more than 20 to 50%. of air flows through the HEPA filter than ULPA filters to maintain optimum cleanliness operator zone. Lower pressure drop or loss and longer filter life than ULPA filters.

ERGONOMIC 10° ANGLE WINDOW design, coupling with adjustable support, stand with telescopic leg/stand with brake casters


UV RESISTANT TEMPER SAFETY GLASS window sash 6 mm thickness.

Advanced Human Machine Interface (HMI) Touchscreen mounted on SAFE ZONE BSCs to provide users with a better graphic view for easy operation and function settings and continuously displays the cabinet’s performance

High-resolution LCD visualizes operating, monitoring and efficient control in real-time. The large 7-inch full-colour touch screen enables intuitive parameter entry and displays variable data, including trend graphs and alarm elements. User-friendly visual icons quickly create users an “easy-to-navigate” to make operations more efficient.

Display real-time clock and date, inflow, downflow, negative plenum pressure, HEPA filter running hrs., and UV lamp running hrs. on the main screen

FEATURES:

  • Automatic Pre-purge/ post-purge
  • Built-in timer for counting up and down, UV light
  • Safety interlock for window and UV light to protect the operator from UV irradiation
  • Audible and visual text alarms for the following abnormal situations:
  • High/Low inflow velocity
  • High/Low downflow velocity
  • Wrong window operation height
  • High negative plenum pressure
  • Usage time of the HEPA filter exceeds the setting hour
  • Usage time of UV lamp exceeds the setting hour
  • Operation records including inflow, downflow and Pressure
  • Multi-language display setting
  • Two kinds of unit display settings for air velocity and pressure

 

INCLUDED ACCESSORIES :

  • Stand with Castor level Stand
  • Arm Rest – 1 pair
  • Electrical outlets – 2 No
  • Service fitting – 1 No
  • UV lamp – 1 No
  • White lamp – 1 No 

 

TECHNICAL SPECIFICATIONS

International Certification

NSF 49, EN 12469, ISO 9001:2015

External Dimensions ( W X D X H)

1334 x 789 x 1537 mm

Internal Dimensions (W X D X H)

1220 x 626 x 709 mm

Cabinet Main Construction

  • Working area: stainless steel 304.
  • External construction: cold-rolled steel with powder coating finish.

Air Flow

70% re-circulated / 30% exhausted

Window Operation Height

25 cm

Avg. Inflow Velocity

0.51 m/s

Avg. Downflow Velocity

0.30 m/s

Exhaust Air Volume With Exhaust Canopy

13 m3 /min

Supply / Exhaust HEPA Filter

99.99% @0.3 micron (Class H14)

Light Intensity

>1000 lux

Sound Level

<63 dBA

UV Lamp

254 nm / FL 30W x 1

Fluorescent Lamp

T5/28W x 2

Window Glass

6 mm tempered glass

Electrical Outlets

North America (UL listed), Europe / worldwide (CE certified)

Room Height Suggested

227 cm at least

Power Consumption

300 W

Net Weight

240 kg

Shipping Weight

280 kg

Shipping Dimensions (W X D X H)

1400 x 850 x 1950 ± 30 mm

Video demo